When completing this module the learner will be able to complete the following:

How to ‘start-up’ a Coriolis system to achieve the best measurement results
  - Verify the mechanical installation / sensor and transmitter
  - Verify all wiring, grounding
  - Apply power and verify the ‘status’ on the transmitter (use transmitter LED, HART communicator, ProLink, or other configuration tools)
  - Verify all calibration parameters (calibration constants using the display, HART communicator, ProLink, or other configuration tools)
  - Configure all out channels to desired units, range,…. (use transmitter display, HART communicator, ProLink, or other configuration tools)
  - Simulate output signals as required or run process to verify output signals
  - Perform sensor ‘zero’ as required by the application